Page 16 50081_Assembly.fm
Operating Environment
2.2.3 Connecting the Titrant Bottle
2. Clip the stop ring into the locating hole on the support rod.
Note: The stop ring may be pre-assembled on the support rod.
3. Compress the clips on each side of the electrode head and slide the
head onto the metal rod. Release the two clips to secure the head
in place.
4. Drop the supplied magnetic stir bar into the titration beaker and place
the beaker on the titration platform.
Note: The provided stir bar has been specifically selected for use with the
AutoCAT 9000. A different stir bar may spin erratically and produce
unwanted turbulence and added noise in the titration curve.
1. Place the titrant bottle into the bottle holder.
2. Thread the desiccant column onto the supplied bottle stopper.
3. Thread the appropriate bottle stopper (for 500 mL or 1000 mL titrant
bottle) onto the new titrant bottle and place it into the holder.