EasyManua.ls Logo

Hach AutoCAT 9000 - Software Setup for Removing or Replacing the Burette

Hach AutoCAT 9000
176 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Software Setup
Software Setup
50081_SoftwareSetup.fm Page 37
4.3.1 Software Setup for Removing or Replacing the Burette
To remove the burette:
1. Press
6 to select Remove Burette.
2. Press the
CHECK MARK to confirm.
3. Follow the embedded procedure (including the purge and rinse cycles).
4. Remove the burette.
To replace the burette:
1. Press
5 to select Replace Burette.
2. Press
1 to confirm.
3. Highlight Burette Volume and press the
CHECK MARK to confirm.
4. Input the replacement burette volume and press the
CHECK MARK to validate.
5. Highlight Serial Number and press the
CHECK MARK to confirm.
6. Input the replacement burette serial number and press the
CHECK MARK to validate.
7. Press
1 to confirm.
8. Remove the burette and press the
CHECK MARK to confirm.
9. Install the replacement burette and press the
CHECK MARK to confirm.
9. Verify the burette
information.
Press
1 to confirm.
10. Install the burette if not
already installed. See
Section 2.2.4 Assembling
the Burette on page 17 for
complete instructions.
Press the
CHECK MARK to
validate.
11. The burette will
automatically reposition
after the install.
12. Press Esc to return to
the Titrants menu. Proceed
to Section 4.4 on page 38.
Note: When the burette
installation is complete the
Replace Burette and
Remove Burette options will be
displayed.

Table of Contents