Basic Operation
Basic Operation
50081_Operation.fm Page 31
Note:Any changes made with the Edition feature will take effect on subsequent procedures— not
on the current sample titration.
3.5 Using Sub-ID's
The AutoCAT 9000 allows the user to store optimized parameters for up to three Sub-IDs
for each method. The user can create these Sub-IDs with parameters optimized for a
specific point of the process they need to monitor.
This eliminates the need for operators to change the titration parameters for samples
from different points of their process: they need only to select an appropriate Sub-ID.
1. To change the Sub-ID in Routine Mode, select Display Method on the Main Menu
screen. In Supervisor Mode, select Method Library on the Main Menu screen.
2. From this screen, both the Method and the Sub-ID can be changed. To change the
Method or the Sub-ID select ID or Sub-ID respectively.
3. Initially these Sub-IDs have identical default parameters, and are simply named
A, B, C. The user can assign names to these Sub-IDs, such as Influent, Effluent, Low,
Medium, High, Tap 1, Tap 2, Pre-Dechlor, etc.
4. The individual parameters for a Sub-ID (Predose, Max. Volume, Increment, and
Current Range) can be changed by selecting any of the titrations listed for the method.
Note: More information on the effects of changing these parameters can be found in
Collecting and Accessing Data (page 43).
5. To return to the Main Menu press ESC.