Software Setup
Software Setup
Page 38
50081_SoftwareSetup.fm
1. From the Titrants menu,
press
2 to select Install
Titrants.
2. Press 1 to select
Titrants Setup.
3. Highlight Lot no.
Press the
CHECK MARK to
input or change.
4. Use the alphanumeric
keypad to input or change
the titrant lot number—
located next to the barcode
on the titrant bottle.
Press the
CHECK MARK to
validate.
5. Highlight Expiration.
Press the
CHECK MARK to
input or change.
6. Use the alphanumeric
keypad to input the month
and year of the titrant
expiration date (located next
to the barcode on the titrant
bottle).
Press the
CHECK MARK to
validate.
7. Verify the titrant
information. Press
1 to
confirm.
8. Press 2 to select
No Rinse.
Note: Initial installation does
not require a burette rinse;
however, it is highly
recommended for titrant
replacement or exchange.
4.4 Software Setup for Titrant Installation