Basic Operation
Basic Operation
50081_Operation.fm Page 29
3.3.7 Printer Status
See Printer Installation and Setup on page 153 for the printer installation procedure.
1. Select the printer status using the
DOWN key to highlight Printer.
Press the
CHECK MARK to confirm the selection.
2. Press the
UP or DOWN key. With the desired setting highlighted, press
the
CHECK MARK to validate and return to the Configuration Menu.
Note:If the printer option is enabled, a printer must be connected to the instrument. If no printer
is connected, an error message will be displayed and the instrument will try to detect a
printer for extended periods of time.
3.3.8 Beep ON/OFF
This selection toggles the ON/OFF status of the alarm beeper. If enabled, the AutoCAT
will emit an audible beep to alert the user to analysis completion, titrant expiration, etc.
1. Highlight Beep.
2. Press the
CHECK MARK to confirm the selection.
3. Press the
UP or DOWN key
4. Press the
CHECK MARK to validate and return to the Configuration Menu.
3.3.9 Rules for Routine Access Users
1. Access the Setup menu.
2. Select Access Routine Mode.
The
Supervisor defines which operations can be ignored in routine mode. Configure
Demand and Alarm parameters for Titrant calibration by highlighting the desired
selection then pressing The configuration options are as follows:
Demand: Unlocked
The routine user can ignore the demand for titrant calibration and continue the
measurements.
Demand: Locked
The routine user must either calibrate the titrant or input a new titrant concentration
before continuing the measurements.
Alarm: Unlocked
The routine user can accept a titrant calibration even if the analysis result lies outside the
acceptance range specified by the method.
Alarm: Locked
The routine user must repeat the titrant calibration when the analysis result lies outside
the acceptance range specified by the method.