Software Setup
Software Setup
Page 40
50081_SoftwareSetup.fm
14. Press 1 to confirm.
15. Select Rinse Before Replacing and follow the embedded method procedure.
1. From Titrants menu,
press
1 to select
Calibrate/Enter
Titrant Conc.
2. Highlight PAO 5.64mN
(or another method) and
press the
CHECK MARK to
view the list of available
titrants for the chosen
procedure.
Note: 5.64 mN is equivalent to
0.00564 N.
3. Highlight the titrant
identifier (PAO 5.64mN in
this example) and press the
CHECK MARK to validate.
Note: The AutoCAT 9000
automatically stores installed
titrants for later concentration
updates or calibrations.
4. Press 1 to select Enter
Titrant Concentration.
5. Highlight User and
press the
CHECK MARK
to input or change the
user ID.
6. Use the alphanumeric
keypad to input the user ID.
Press the
CHECK MARK to
validate.
Note: The user ID may be any
combination of alphanumeric
characters up to 16 characters
in length.
7. Verify the user ID and
press
1 to confirm.
8. Highlight Titrant Conc.
and press the
CHECK MARK
to input or change the
concentration.
4.5 Setting Titrant Concentration