PAO/5.64 mN
PAO/5.64 mN
Page 52
50081_PAONaThioCalibration.fm
1. Turn the instrument on
if necessary and access the
Titrants menu.
Press
4 to select Titrant
Library.
Note: Supervisor-mode is
required to access calibrations.
See Accessing Supervisor
Mode on page 33.
2. Highlight ID.
Press the
CHECK MARK
to change.
3. Highlight
PAO 5.64mN.
Press the
CHECK MARK
to validate.
Note: PAO 5.64mN has been
chosen for demonstration
purposes only.
4. Press 1 to select
Edit titrant.
5. Press the CHECK MARK
to change the titrant
concentration input mode.
6. Highlight Calibrate.
Press the
CHECK MARK
to validate.
7. The titrant
concentration input mode is
now set to Calibrate and
calibration parameters are
available by scrolling this
screen. Press
ESC twice to
return to the Titrants menu.
8. Press 1 to select
Calibrate/Enter titrant conc.
Note: Titrant calibration is
required before further
analyses may be run, as
indicated by the “Stormy” icon.
AutoCAT Calibration Method PAO/5.64 mN Titrant