Chapter 6, Quality Control 73
Haemonetics
®
TEG
®
6s User Manual P/N 115191-IE, Manual revision: AB
Figure 6-4, Test information screen
The Preparing Test – Load sample screen displays.
Figure 6-5, Preparing Test – Load sample screen
6. Prepare the QC material according to the instructions in the package
insert.
Note: Each QC vial, once reconstituted, contains 1ml of QC material,
which is sufficient for two QC assays.
7. Pipette the prepared QC sample into the cartridge sample port, filling up
to or above the line marked on the cartridge.