The “ ” along with “STIR” tag will be displayed and “WAIT” will blink on the LCD
until the reading is stable.
When the reading is stable and close to the selected buer, “CFM” tag will
blink. Press CFM to conrm calibration.
Repeat procedure with additional pH buers. A total of ve pH buers can be
utilized.
After conrming the last desired buer calibration
points, press CAL or if all ve buer values were
calibrated the instrument will automatically display
“SAVING”, as it stores information and it will then
return to normal measurement mode.
Each time a buer is conrmed, the new calibration
data replaces the old data for the corresponding
buer or for any near buer in a ±0.2 pH proximity.
If current buer has no previous data stored and the
calibration is not full (ve buers), the current buer
is added to the existing calibration. If the existing
calibration is full, the instrument asks which buer
to replace.
Press the ARROW keys to select another buer to be replaced.
Press CFM to conrm the buer that will be replaced.
Press CAL to leave calibration without replacing.
Note: If the replaced buer is outside the ±0.2 pH window around each of the
calibrated buers, it is possible to select this buer for next calibration during
current calibration.
pH
Calibration