Operational Guide
33
Working With Custom Buers
If a custom buer was set in SETUP menu, it can be selected during calibration
by pressing the ARROW keys. The “C1” or “C2” tag will be on once selected.
Press u if you want to modify the custom buer
value. The buer value will start blinking. Use the
ARROW keys to change the buer value.
After 5 seconds, the buer value is updated. Press u
if you want to change it again.
Note: Custom buer value can be adjusted in a ±1.00 pH window around the
set value during calibration. When a custom buer is displayed, the “C1” or
“C2” tag is displayed.
First Calibration Point
When performing a new calibration, or adding to an existing one, the user
has a choice how the rst new calibration point will be treated in reference
to the existing calibration point. This is selected in SETUP by the parameter
FIRST CALIBRATION POINT. The two SETUP selectable options are “POINT” or
“OFFSET”.
Point: A new buer can be added to an existing calibration. The electrode
slope will be reevaluated with the addition of this buer (normal operation).
Oset: The new buer calibration point can create a constant oset to all
existing pH calibration data (existing calibration must have a minimum of two
pH buers).
pH
Calibration