CALIBRATION PROCEDURE
Note: If you’re using different calibration standards, change the displayed value by pressing UP
or DOWN keys until the display shows the desired value.
Remove the <0,10 NTU standard cuvette and place the 15.0 NTU standard cuvette into the holder.
Close the lid and press READ “----” and the “Lamp and Read Status” indicator will blink again
while making the reading.
At the end of the reading, the third calibration point (100 NTU) and “CAL P.3” is displayed. If
desired, the value can be changed using the UP or DOWN keys.
Note: Press CAL/RCL to exit calibration. The instrument will memorize the two-point (<0.10 and
15.0 NTU) calibration data and will return to measurement mode.
Remove the 15.0 NTU standard cuvette from the meter and place the 100 NTU standard cuvette
into the holder.
Close the lid and press READ “----” and the “Lamp and Read Status” indicator will
blink again while making the reading. At the end of the reading, the fourth calibration point