The instrument will display a lot of data including calibration standards, offset, time and date, etc.
Use the / keys to select the offset or calibration standards, in order to view new information.
To see more information press More.
More information regarding offset.
More information regarding standards.
Notes: The standards that are displayed in video inverse mode are from previous calibrations.
“No user calibration” message is displayed if all calibrations are cleared or the instrument was
not calibrated in the EC range.
LAST NaCl % CALIBRATION DATA
Last NaCl calibration data is stored automatically after a successful calibration.
To view the NaCl calibration data, press GLP key while in NaCl measurement mode.
The instrument will display the NaCl information: calibration date, time and offset.