CHR xx Change the instrument range according with the parameter value (xx):
• xx=10 EC range
• xx=11 Resistivity range
• xx=12 TDS range
• xx=13 USP range
• xx=14 NaCl % range
• xx=15 Salinity, Seawater range
• xx=16 Salinity, PSU range
The instrument will answer for these commands with:
<STX> <answer> <ETX>
where: <STX> is 02 ASCII code character (start of text)
<ETX> is 03 ASCII code character (end of text)
<answer>:
<ACK> is 06 ASCII code character (recognized command)
<NAK> is 21 ASCII code character (unrecognized command)
<CAN> is 24 ASCII code character (corrupted command)
COMMANDS REQUIRING AN ANSWER
The instrument will answer for these commands with:
<STX> <answer> <checksum> <ETX>
where the checksum is the bytes sum of the answer string sent as 2 ASCII characters.
All the answer messages are with ASCII characters.
RAS Causes the instrument to send a complete set of readings in according with the current
range:
EC and temperature readings or EC range.
Resistivity, EC and temperature readings on Resistivity
TDS, EC and temperature readings on TDS range
Salinity, EC, and temperature or Salinity range.
The answer string contains:
Meter mode (2 chars):
xx=10 EC range
xx=11 Resistivity range
xx=12 TDS range
xx=13 USP range
xx=14 NaCl % range
PC INTERFACE

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI98192 and is the answer not in the manual?

Hanna Instruments HI98192 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI98192
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals