• Press CFM to confirm.
The second expected calibrated point is displayed.
Immerse the probe into the second vessel as close as possible to the reference thermometer. Allow
a few seconds for the probe to stabilize.
Use the / keys to set the calibration point value to that measured by the reference
thermometer.
When the reading is stable and within range of the selected calibration point, CFM functional key
is displayed.
• Press CFM to confirm. The instrument returns to measurement mode.
TEMPERATURE CALIBRATION (for technical personnel only)