5 Mechanical installation
5.1 Important measures for installation
Treat the transducer gently
Use appropriate lifting tools
The mounting surfaces (substructure and pressure plates) must
− maintain an evenness of 0.02 mm under all loading
− be free from grease and dust
Do not overload the transducer, even temporarily e.g. through unevenly
distributed bearing loads
During straightening work insert supporting bodies of the same height
(dummies) first if necessary.
WARNING
If there is a risk of breakage through overload on the transducer and
thus a risk to persons, additional safety measures are to be taken.