C6A
18
A0638−2.3 en/de/frHBM
6 Connections
6.1 Instructions for cabling
Electrical and magnetic fields often cause the induction of noise/interference
voltage in the measuring circuit. This interference is primarily caused by
high−power transmission lines parallel to the measuring leads, but also by
contactors and electric motors in the vicinity.
Please note the following instructions:
Always use shielded, low−capacity measurement cable (HBM cables meet
these requirements).
Do not lay measurement cable parallel to high−voltage power lines or
control circuits. If this is not possible (e.g. in cable ducts) protect the
measurement cable, e.g. with armoured steel tube and maintain a minimum
distance of 50 cm from the other cables. High−voltage power lines and
control lines should be twisted (15 turns per metre).
Avoid stray fields of transformers, motors and contactors.
Do not earth transducer, amplifier and display device more than once. All
the devices in the measuring chain are to be connected to the same
earthed conductor.
The screen of the connection cable is connected to the transducer housing.