32
CMD600
A28201.1 en
DATA_STREAM_TARGET
Set /get IP address and UDP port, to which real time data will be send
command DATA_STREAM_TARGET IP_address, UDP_port
parameters:
IP_address string Format “XXX.XXX.XXX.XXX” Default value
“0.0.0.0”, which is not valid IP. Error will be
returned, if data stream is enabled with this IP.
UDP_port int Default value 12345.
response OK, DATA_STREAM_TARGET = IP_address,UDP_port
parameters:
IP_address string See description above
UDP_port int See description above
inquiry DATA_STREAM_TARGET = ?
response OK, DATA_STREAM_TARGET = IP_address,UDP_port
IP_address string See description above
UDP_port int See description above
help DATA_STREAM_TARGET?
response OK, DATA_STREAM_TARGET IP_address, UDP_port
DATA_STREAM_RATE
Set /get rate at which real time data is send
command DATA_STREAM_RATE rate
parameters:
rate float Rate (frequency) at which real time data is
send. Minimum 1, maximum 1000. Default 1.
response OK, DATA_STREAM_RATE = actual_rate
parameters:
actual_rate float Rate (frequency) at which real time data is
send. This is real value decimated to resolution
of internal timers in charge amplifier and is not
necessary full equal to requested value
inquiry DATA_STREAM_RATE = ?
response OK, DATA_STREAM_RATE= actual_rate
parameters:
actual_rate float See description above.
help DATA_STREAM_RATE?