35
CMD600
A28201.1 en
RESTART
Perform full reset to charge amplifier. This is same as cycling power supply. Used to
activate new network settings. Amplifier will be restarted 1 second after transmitting
response to this command.
command RESTART
response OK, RESTART
inquiry Not applicable
help RESTART?
response OK, RESTART used to restart amplifier
FACTORY_DEFAULT
This command set both packets of parameters to default values. If protection of
parameters is disabled, then all parameters set to factory default, except
ETHERNET_ADDRESS and protection parameter (PROTECT_PARAM). If protection of
parameters is enabled, then all unprotected parameters set to factory default. This
includes also IP settings. Amplifiers restore and restart will be done 1 second after
transmitting response to this command. Communication with amplifier will be probably
lost.
command FACTORY_DEFAULT
response OK, FACTORY_DEFAULT
inquiry Not applicable
help FACTORY_DEFAULT?
response OK, FACTORY_DEFAULT restore to factory default and restart
Factory default can be triggered also with SensorTeach button. See “Factory default
procedure”.
HELP
command HELP
response OK,HELP