CMD600
A28201.1 en
8 System parameters
There are 2 sets of parameters. First set starts on address 0, second set
starts on address 2400. Both sets have the same structure of parameters.
Some parameters are the same in both packets. Those parameters are global
parameters and when parameter is changed then is changed also parameter
on inactive packet.
Below is a list of all parameters, which are stored in EEPROM of charge
amplifier. Protected parameters are written during production and calibration.
In normal use should not be accessed, because complete functionality is
available via command interface.
Num-
ber
Protecti
on
Glo-
bal
Default value Description
0 Y Y 300 Input offset value in quants. Used to
correct input operational amplifier offset
during drift compensation
(min = 2048, max= 2047).
1 Y Y +4.9762E3 DAC_k
2 Y Y 1.2002E+01 With parameters 1and 2 (DAC_n) is
calculated DAC value in quants, for
desired output voltage: DAC_value =
(output_valueDAC_n)/DAC_k
3 Y Y 4.3737E4 Corr_offset(k)
4 Y Y 2.2000E3 With parameters 3 and 4 Corr_offset(n)
is calculated correction offset.
Corr_offset =
Corr_offset(k)*Actualy_PGA_gai n+
Corr_offset(n).
5 Y Y 1.0000E4 LPF_k
6 Y Y 1.8000E2 With parameters 5 and 6 (LPF_n) is
calculated correction offset when LPF is
enabled:
LPF_corr_offset =
LPF_k*Actualy_PGA_gain+ LPF_n
7 N N 0.0000E+00 Offset value[V](min = 10,max=10)
8 Y Y 4.8447E3 AD_k
9 Y Y 3.7000E2 With parameters 8 and 9(AD_n) is
calculated output voltage:
Output_voltage =
AD_k*AD_quants+AD_n
10 N N N Engineering unit (max 5 characters).
11 N N 1.0000E+00 Sensor sensitivity [V/EnUn]
12 Y Y 0 Protection parameter enabled
(0..protection disabled,1..protection
enabled)