47
Error Code Description Corrective Actions
E025
PT result is above
reportable range.
Please retest with a new test
strip. If the problem recurs,
please contact Heska's
Technical Support Services and
use the laboratory method to
perform the test.
E026
PT result is below
reportable range.
Please retest with a new test
strip. If the problem recurs,
please contact Heska's
Technical Support Services and
use the laboratory method to
perform the test.
E030
The analyzer is
vibrating or shaking.
Do not move, vibrate, or shake
the analyzer during the test and
ensure that the analyzer is used
in an environment away from
high frequency vibrations.
E033
Codechip reading
error.
Please unplug the Codechip and
reinsert it, and make sure the
Codechip is inserted in place.
If the problem recurs, please
contact Heska's Technical
Support Services.
NO COAG
No sample coagulation
of PT is detected.
Please retest with a new test
strip. If the problem recurs,
please contact Heska's
Technical Support Services and
use the laboratory method to
perform the test.