EasyManua.ls Logo

Hettich EBA 200 - Querying Operating Hours and Centrifugation Runs

Hettich EBA 200
45 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
AB1800en_SARev.: 03 / 11.2023
28 / 39
Software operation
Resetting the cycle counter
The centrifuge is equipped with a cycle counter. The cycle counter counts
the run cycles (centrifugation cycles). The remaining number of run cycles
(centrifugation runs) is displayed briey after each centrifugation run.
If the maximum permissible number of rotor run cycles entered is
exceeded,
‘Cycles passed’
is displayed after each start of a centrifugation
run. The centrifugation run must be restarted. The rotor must be replaced
with a new one.
Once the rotor has been replaced, the cycle counter must be reset to
‘0’
.
The cycle counter must be reset to
‘0’
after installing a new rotor.
1. Press and hold the
[SELECT]
button.
🡆
‘*MACHINE MENU*’
is displayed after 8 seconds.
2. Press the
[SELECT]
button repeatedly until
‘-> Time & Cycles’
is
displayed.
3. Press the
[START/PULSE]
button.
4. Press the
[SELECT]
button repeatedly until
‘Cyc sum=...’
is displayed.
5. Press the
[RCF]
button.
6.
Press the
[t
]
button.
🡆 The number of run cycles completed is reset to
‘0’
.
7. Press the
[START/PULSE]
button.
🡆
‘Store cycles...’
is displayed.
8. Press the
[STOP/OPEN]
button twice to exit the
‘-> Time & Cycles’
menu
or
Press the
[STOP/OPEN]
button three times to exit the
‘*MACHINE
MENU*’
.
7.2.3 Querying operating hours and centrifugation runs
The operating hours are divided into internal and external operating hours.
Internal operating hours: Total time for which the device has been
switched on.
External operating hours: Total time of centrifugation runs to date.
The rotor is stationary.
1. Press and hold the
[SELECT]
button.
🡆
‘*MACHINE MENU*’
is displayed after 8 seconds.
2. Press the
[SELECT]
button repeatedly until
‘-> Time & Cycles’
is
displayed.
3. Press the
[START/PULSE]
button.
🡆
‘TimeExt=’
is displayed.
TimeExt: External operating hours
4. Press the
[SELECT]
button.
🡆
‘TimeInt=’
is displayed.
TimeInt: Internal operating hours
5. Press the
[SELECT]
button.
🡆
‘Starts=’
is displayed.
Starts: Number of all centrifugation runs

Table of Contents

Other manuals for Hettich EBA 200

Related product manuals