Troubleshooting
Fault description Cause Remedy
COM ERROR 31-36 Error/defect in electronics.
Perform a MAINS RESET.
FC ERROR 60, 61.1-61.21,
61.64-61.142
Error/defect in electronics.
Perform a MAINS RESET.
FC ERROR 61.23 Speed measurement error.
Do not switch off the device while
‘Rotation’
is displayed.
Perform a MAINS RESET if
‘Lid
locked’
is displayed.
TACHO ERR 61.22 Speed measurement error.
Do not switch off the device while
‘Rotation’
is displayed.
Perform a MAINS RESET if
‘Lid
locked’
is displayed.
FC ERROR 61.153 Error/defect in electronics.
Perform a MAINS RESET.
Open the lid.
Check the loading of the rotor.
Repeat the centrifugation run.
The left
half of the display lights up.
-
Notify customer service.
9.2
Perform a MAINS RESET
1.
Set the mains switch to
[0]
.
2.
Wait 10 seconds.
3.
Set the mains switch to
[I]
.
9.3
Emergency release
The lid cannot be unlocked by the motor in the event of a power failure.
Emergency unlocking by hand must be performed.
WARNING
Risk of electric shock due to maintenance and servicing
work on live device.
Disconnect the device from the mains before carrying
out repairs and maintenance.
WARNING
Danger of cutting and crushing due to moving rotor.
Do not open the lid until the rotor has stopped.
Personnel:
Trained user