NOTICE
Damage to the device and samples due to premature pro-
gram termination.
Premature program termination is caused by power failure,
switching off during the program or pulling out the mains
plug.
− Do not switch off the device while the program is run-
ning.
−
Do not trigger the emergency release on the device
while the program is running.
− Do not pull out the mains plug while the program is
running.
3 Device overview
3.1 Technical data
Manufacturer Andreas Hettich GmbH & Co.KG,
D-78532 Tuttlingen
Model EBA 280 EBA 280 S
Type 1101 1101-01 1102 1102-01
Mains voltage (±10%) 200-240V1~ 100-127V1~ 200-240V1~ 100-127V1~
Mains frequency 50-60 Hz 50-60 Hz 50-60 Hz 50-60 Hz
power consumption 185 VA 185 VA 330 VA 330 VA
Power consumption 0.85A 1.75A 1.6A 3.0A
max. capacity 6 x 50 ml
max. permissible density 1.2kg/dm³
max. speed (RPM) 6000
max. acceleration (RCF) 4146 5071
max. kinetic energy 2700 Nm
Obligation to perform
checks
(DGUV Rules 100-500)
(valid only in Germany)
No
Ambient conditions (EN / IEC 61010-1):
Installation site indoors only
Altitude up to 2000 m above sea level
Ambient temperature 2°C to 40 °C