AB1101en_SARev.: 03 / 11.2023
7. Press the
[STOP/OPEN]
button once to exit the
‘-> Settings’
menu
or
Press the
[STOP/OPEN]
button twice to exit the
‘*MACHINE MENU*’
.
7.4.8
Centrifugation data displayed after switching on
The centrifugation data of program 1 or the last program used is displayed
after switching on.
1. Press and hold the
[SELECT]
button.
🡆
‘*MACHINE MENU*’
is displayed after 8 seconds.
2. Press the
[SELECT]
button repeatedly until
‘-> Settings’
is displayed.
3. Press the
[START/PULSE]
button.
🡆
‘End beep = on’
or
‘End beep = off’
is displayed.
4. Press the
[SELECT]
button repeatedly until
‘Start Pr = First’
or
‘Start
Pr = Last’
is displayed.
5. Use
[Adjustment buttons]
to set
‘Last’
or
‘First’
.
Last = last program used
First = program 1
6. Press the
[START/PULSE]
button.
🡆 The setting is stored.
‘Store setting...’
is displayed briey.
‘-> Settings’
is then displayed.
7. Press the
[STOP/OPEN]
button once to exit the
‘-> Settings’
menu
or
Press the
[STOP/OPEN]
button twice to exit the
‘*MACHINE MENU*’
.
7.4.9
Indicator backlight
The indicator backlight can be switched off after 2 minutes to save energy.
The rotor is stationary.
1. Press and hold the
[SELECT]
button.
🡆
‘*MACHINE MENU*’
is displayed after 8 seconds.
2. Press the
[SELECT]
button repeatedly until
‘-> Settings’
is displayed.
3. Press the
[START/PULSE]
button.
🡆
‘End beep = on’
or
‘End beep = off’
is displayed.
4. Press the
[SELECT]
button repeatedly until
‘Power save=off’
or
‘Power save=on’
is displayed.
5. Use
[Adjustment buttons]
to set
‘off’
or
‘on’
.
off: Backlight is switched off.
on: Backlight is switched on.
6. Press the
[START/PULSE]
button.
🡆 The setting is stored.
‘Store setting...’
is displayed briey.
‘-> Settings’
is then displayed.