AB1101en_SARev.: 03 / 11.2023
36 / 44
Cleaning and care
Cleaning the device
Cleaning the accessories
The device and its accessories must not be cleaned in dishwashers.
Only perform hand cleaning and liquid disinfection.
The water temperature must not exceed 25°C.
To prevent any corrosion due to use of detergents or disinfectants, it is
essential to follow the special application instructions provided by the
manufacturers of the detergent or disinfectant.
Disinfectant:
Surface disinfectant (not disinfectant for hands or instruments)
Ethanol as the sole active substance.
Do not use an ethanol-propanol mixture to disinfect the viewing window
in the lid of the device.
Concentration is not less than 30%
pH: 6 – 8
Non-corrosive
8.3 Cleaning
1. Open the lid.
2.
Switch off the device and disconnect it from the power supply.
3.
Remove accessories.
4.
Clean the centrifuge housing and the centrifuging chamber with soap
or a mild detergent and a damp cloth.
5.
Remove any detergent residues with a damp cloth after using deter-
gents.
6.
The surfaces must be dried immediately after cleaning.
7.
Dry the centrifuging chamber with an absorbent cloth if condensation
forms.
1.
Clean the accessories using the detergent and a damp cloth.
2.
Remove any detergent residues with a damp cloth after using deter-
gents.
3.
Dry the accessories immediately after cleaning using a lint-free cloth
and oil-free compressed air. Dry all cavities completely using oil-free
compressed air.
8.4
Disinfection
Disinfection must always be preceded by cleaning of the
components concerned.
See
Chapter 8 ‘Cleaning and care’ on page 35
Disinfectant concentration and application time according
to the manufacturer's instructions.