EasyManua.ls Logo

Hettich EBA 280 - Switching the Centrifuge On and Off; Switching the Centrifuge On; Switching Off the Centrifuge; Operation

Hettich EBA 280
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19 / 44
Rev.: 03 / 11.2023
AB1101en_SA
Operation
Switching the centrifuge on
Switching off the centrifuge
Opening the lid
Closing the lid
5.4 Switching the centrifuge on and off.
Personnel:
Trained user
Set the mains switch to
[I]
.
🡆 The buttons ash, depending on the centrifuge type.
The following indicators appear one after the other, depending on
the centrifuge type:
the centrifuge model
the machine type and program version
the rotor code (R) and the maximum speed (maxRPM) of the
last rotor recognised by the rotor detection
When the lid is closed: For centrifuges with cooling,
‘Open the
lid’
indicator. The lid opens in centrifuges without cooling.
When the lid is open: The centrifugation data of the last pro-
gram used or program 1.
The rotor is stationary.
Set the mains switch to
[0]
.
6
Operation
6.1
Opening and closing the lid
Personnel:
Trained user
The centrifuge is switched on.
The rotor is stationary.
Press the
[STOP/OPEN]
button.
🡆 The lid unlocks by means of a motor.
The
‘Lid unlocked’
indicator appears.
CAUTION
Danger of crushing when closing the lid.
Danger of crushing ngers when the closing motor pulls the
lid against the seal.
No parts of the body should be in the hazard zone of the
lid when closing the lid.
To close the lid, press on the lid from above.
NOTICE
Damage to the device caused by the lid slamming.
Close the lid slowly.
Do not slam the lid.
Personnel:
Trained user