Brake level
Changing the runtime
Start of runtime count
3. Press the
[START/PULSE]
button.
🡆 The setting is transferred to the indicator.
1. Press the
[SELECT]
button repeatedly until is displayed.
2. Use the
[Adjustment buttons]
to set the desired value.
A numerical value can be set from 0 to 9.
Adjustable in 1 second increments.
9 = shortest ramp-down time
1 = long ramp-down time
0 = longest ramp-down time (unbraked ramp-down).
3. Press the
[START/PULSE]
button.
🡆 The setting is transferred to the indicator.
7.1.2 Runtime
For continuous operation, the minutes and seconds must
be set to zero.
1.
Press the
[SELECT]
button repeatedly until
‘t/min’
is displayed.
2.
Use the
[Adjustment buttons]
to set the desired value.
A numerical value can be set from 1 to 99 minutes.
Adjustable in 1 minute increments.
3.
Press the
[SELECT]
button.
🡆
‘t/sec’
is displayed.
4.
Use the
[Adjustment buttons]
to set the desired value.
A numerical value can be set from 1 to 59 seconds.
Adjustable in 1 second increments.
5.
Press the
[START/PULSE]
button.
🡆
The settings are shown in the indicator.
■ The
‘Dual time’
function is enabled. The function is enabled ex works.
1.
Press the
[SELECT]
button repeatedly until
‘Begins at START’
or
‘Begins at SPEED’
is displayed.
2.
Use the
[Adjustment buttons]
to select the desired setting.
■
‘Begins at START’
= Runtime counting begins after the start of the
centrifugation run.
■
‘Begins at SPEED’
= Timing of the runtime starts after the set
speed is reached.
This is indicated by the
‘é’
symbol in the indicator to the left of the
time.
3.
Press the
[START/PULSE]
button.
🡆 The settings are shown in the indicator.
7.1.3 Speed, RPM
1.
Press the
[SELECT]
button repeatedly until
‘RPM’
is displayed.