AB1101en_SARev.: 03 / 11.2023
7.2.2 Entering or changing programs
Program locations 1-9 are available.
No programs may be stored in program location #. Pro-
gram location # serves as a buffer for changed centrifuga-
tion parameters.
A dash "-" appears in the indicator instead of the program
location number if centrifugation parameters are changed
but then not saved to a program location. The centrifuga-
tion parameters are automatically saved in program location
# after starting the centrifugation run.
The centrifugation parameters in program location # are
overwritten every time a centrifugation run is executed
with changed centrifugation parameters that have not been
saved to a program location.
1. If required: Press the
[RCF]
button to toggle between RPM and RCF
indicator.
🡆
The indicator above the button lights up.
2.
If required: Press the
[SELECT]
button to select the desired parameter
and set it with the
[Adjustment buttons]
.
The parameters t/min and t/sec must be set to 0 with the
[Adjustment
buttons]
to set continuous operation. Continuous operation is shown
in the time indicator with
‘--:--’
.
3. Use the
[SELECT]
button to select the
‘PROG STO’
parameter.
4. Use the
[Adjustment buttons]
to set the desired program location.
5. Press the
[START/PULSE]
button.
🡆 Settings are stored in the desired program location.
‘*** OK ***’
is displayed briey.
The settings are always stored in program location # if the
[START/
PULSE]
button is pressed without the
‘PROG STO’
parameter
being selected.
7.2.3
Saving programs
1.
Press the
[PROG]
button twice.
🡆
‘PROG STO’
is displayed.
PROG STO: Program location where the centrifugation parameters
are saved.
2. Use the
[Adjustment buttons]
to set the desired program location.
3. Press the
[START/PULSE]
button.
🡆 Settings are stored in the desired program location.
‘*** OK ***’
is displayed briey.
The settings are always stored in program location # if the
[START/
PULSE]
button is pressed without the
‘PROG STO’
parameter
being selected.
7.2.4
Automatic buffer
The centrifugation data is saved to program location
‘#’
every time a centri-
fugation run is started.
No programs can be stored in program location
‘#’
.