7. Press the
[STOP/OPEN]
button once to exit the
‘-> Settings’
menu
or
Press the
[STOP/OPEN]
button twice to exit the
‘*MACHINE MENU*’
.
8 Cleaning and care
8.1
Overview table
Chap. Task to execute
8 Cleaning and care 35
8.3 Cleaning 36
8.3 Cleaning the device X 36
8.3 Cleaning the accessories X 36
8.4 Disinfection 36
8.4 Disinfecting the device X 37
8.4 Disinfecting the accessories X 37
8.5 Maintenance 37
8.5 Greasing the rubber seal of the centrifuging
chamber
X 37
8.5 Trunnion greasing X 37
8.5 Checking the accessories X 37
8.5 Inspecting the centrifuging chamber for damage X 38
8.5 Greasing the motor shaft X 38
8.5 Accessories with a limited service life X 38
8.5 Replacing centrifuge tubes X 38
8.2
Cleaning and disinfection instructions
DANGER
Risk of contamination for the user due to inadequate
cleaning or failure to observe the cleaning instructions.
− Observe cleaning instructions.
− Wear personal protective equipment when cleaning the
device.
− Observe laboratory regulations (e.g. TRBAs, the German
Protection against Infection Act, hygiene plan) for han-
dling biological agents.