Fig. 17: Transport lock
1 Spacer sleeve
2 Screw
Setting up the centrifuge
2. Remove the padding.
3. Remove the device and accessories by lifting them up out of the box.
4. Place the device on a stable and level surface.
5.2 Remove the transport lock
Personnel:
■
Trained user
The lid is closed.
The mains cable is disconnected from the device.
1. Tilt the device on the back of the device.
2. Unscrew 2 screws (
2
).
3. Remove 2 spacer sleeves (
1
).
4. Keep the screws and spacer sleeves in a safe place.
5.3
Setting up and connecting the centrifuge
WARNING
Risk of injury due to failing to maintain a
sufcient distance to
the centrifuge.
− As per EN/IEC 61010-2-020, no persons, hazardous
materials or objects may be present within a safety zone
of 300mm around the centrifuge during a centrifugation
run.
−
A distance of 300mm from the ventilation slots and ven-
tilation openings of the centrifuge must be maintained.
CAUTION
Risk of crushing and damage to the device due to it falling
down because of vibration-induced position alterations.
− Place the device on a stable and level surface.
− Select the installation surface dependent on the weight
of the device.