Cleaning and care
Disinfecting the device
Disinfecting the accessories
Autoclaving
Greasing the rubber seal of
the centrifuging chamber
Trunnion greasing
Checking the accessories
CAUTION
Risk of injury due to ingress of water or other liquids.
Protect the device against external liquids.
Do not disinfect the device using spray.
1. Open the lid.
2. Switch off the device and disconnect it from the power supply.
3. Remove accessories.
4. Clean the housing and centrifuging chamber using disinfectant.
5. Remove any disinfectant residues with a damp cloth after using disin-
fectants.
6. The surfaces must be dried immediately after cleaning.
1. Disinfect the accessories using the disinfectant.
2. Wet all cavities with bubble-free disinfectant.
3. Remove the disinfectant residues or leave them to dry after using
disinfectants.
The following accessories may be autoclaved at 121°C / 250°F (20min):
Swing-out rotors
Aluminium angle rotors
Metal buckets
Lid with bioseal
Inserting
No statement can be made about the resulting degree of sterility.
The lids of the rotors and bucket must be removed before autoclaving.
Autoclaving accelerates the ageing of materials. It may cause changes to
colours. After autoclaving, the rotors and accessories are to be visually
inspected for damage and any damaged parts are to be replaced immedi-
ately.
The sealing ring in question is to be replaced if there are signs of cracking,
embrittlement or wear. For lids with non-replaceable sealing rings, the whole
lid must be replaced.
8.5
Maintenance
Rub the sealing ring lightly with a rubber care product.
1.
Remove accessories.
2.
Clean the trunnions.
3.
Remove any detergent residues with a damp cloth after using deter-
gents.
4.
Grease the trunnions and suspension with Hettich Tubenfett 4051.
5.
Excess grease in the centrifuging chamber must be removed.
1.
The accessories are to be checked for wear and corrosion damage.
2.
Check that the rotor is rmly seated.