EasyManua.ls Logo

Hettich EBA 280 - Page 41

Hettich EBA 280
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
39 / 44
Rev.: 03 / 11.2023
AB1101en_SA
Troubleshooting
Fault description Cause Remedy
RPM > ROTOR MAX Speed in the selected program
greater than the maximum
rotor speed.
Check and correct the speed.
MAINS INTERRUPT Loss of mains power during
the centrifugation run. The
centrifugation run was not
completed.
Open the lid.
Press the
[START/PULSE]
button.
If required: Repeat the centrifugation
run.
R... WRONG ROTOR (R =
rotor code)
The rotor being used is not
approved for the device.
Install a rotor that is approved for the
device.
KEYBOARD-ERROR Error/defect in electronics.
Perform a MAINS RESET.
TACHO - ERROR 1, 2, 96.1 Speed pulse failure. No rotor
inserted. Electronics defective.
Perform a MAINS RESET after the dis-
played waiting time (150 seconds) has
elapsed.
Check whether the rotor is inserted.
LID ERROR 4.1-4.127 Lid lock error.
Perform a MAINS RESET.
OVER SPEED 5 Overspeed.
Perform a MAINS RESET.
ROTOR ERROR 10.1-10.6 Rotor coding error.
Perform a MAINS RESET.
VERSION ERROR 12 Wrong centrifuge model
detected. Error / defect in the
electronics.
Perform a MAINS RESET.
UNDER SPEED 13 Underspeed.
Perform a MAINS RESET.
CTRL ERROR 22-25.4 Error/defect in electronics.
Perform a MAINS RESET.
CRC ERROR 27, 27.1 Error/defect in electronics.
Perform a MAINS RESET.
COM ERROR 31-36 Error/defect in electronics.
Perform a MAINS RESET.
FC ERROR 60, 61.1-61.21,
61.64-61.142
Error/defect in electronics.
Perform a MAINS RESET.
TACHO ERR 61.22 Speed measurement error.
The device must not be switched off
while the message "Wait" is displayed.
Perform a MAINS RESET when the
message "wait ...!" is no longer dis-
played.
FC ERROR 61.23 Speed measurement error.
The device must not be switched off
while the message "Wait" is displayed.
Perform a MAINS RESET when the
message "wait ...!" is no longer dis-
played.
FC ERROR 61.153 Error/defect in electronics.
Perform a MAINS RESET.
Check the loading of the rotor.
Repeat the centrifugation run.

Table of Contents