Alarms
Chapter
2:
Troubleshooting
Procedures
For
an
explanation
of
temperature
alarms,
refer
to
table
2-2
on
page
2-18.
Table
2-2.
Temperature
Alarms
Alarm
Description
Low
air
temperature
This
message
appears
with
an
audible
alarm
and
flashing
alarm
indicators
to
indicate
that
the
incubator
air
temperature
is
2.5°C
below
the
air
set
temperature.
To
silence
this
alarm
for
15
min,
press
the
Silence/Reset
key.
Go
to
RAP
2.3.
High
air
temperature
This
message
appears
with
an
audible
alarm
and
flashing
alarm
indicators
to
indicate
that
the
incubator
air
temperature
is
1.5°C
above
the
air
set
temperature.
Additionally,
the
heater
is
turned
off.
To
silence
this
alarm
for
15
min,
press
the
Silence/Reset
key.
Skin
mode
disabled
This
message
appears
with
an
audible
alarm
and
flashing
alarm
indicators
to
indicate
that
Skin
Mode
has
been
disabled.
Activate
Skin
Mode.
Low
skin
temperature
This
message
appears
with
an
audible
alarm
and
flashing
alarm
indicators
to
indicate
that
the
baby’s
skin
temperature
being
monitored
by
skin
temperature
probe
#1
is
1°C
(or
0.5°C)
below
the
skin
set
temperature.
To
silence
this
alarm
for
15
min,
press
the
Silence/Reset
key.
Go
to
RAP
2.3.
High
skin
temperature
High
skin
1
temperature
This
message
appears
with
an
audible
alarm
and
flashing
alarm
indicators
to
indicate
that the
baby’s
skin
temperature
being
monitored
by
skin
temperature
probe
#1
is
1°C
(or
0.5°C)
above
the
skin
set
temperature.
To
silence
this
alarm
for
15
min,
press
the
Silence/Reset
key.
High
temperature
cut
out
This
message
appears
with
an
audible
alarm
and
flashing
alarm
indicators
to
indicate
that
incubator
air
temperature
has
reached
approximately
103.1°F
(39.5°C)
(American
and
Canadian
incuba-
tors
only).
All
other
incubators
will
alarm
at
99.5°F
(37.5°C)
for
set
temperatures
<
99°F
(37°C)
and
103.1°F
(39.5°C)
for
set
temperatures
>
99°F
(37°C).
Page
2
-
18
Isolette®
Infant
Incubator
Service
Manual