6. Install two new O-ring gaskets (P/N: 430-B081-0H0) in the compartment’s inlet and
outlet.
7. Press a new pump assembly (P/N: W01-3002-000) into place, making sure that both the
inlet and outlet from the pump go through the O-rings into the two holes.
8. Insert and tighten the two screws that attach the bracket to the housing.
9. Replace the cover.
10. Turn on the instrument and check for proper pump operation.