EasyManua.ls Logo

Honeywell AreaRAE Plus - Deactivating Policy Enforcement

Honeywell AreaRAE Plus
145 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Deactivating Policy Enforcement
If the AreaRAE Plus/Pro screen displays the message that it must be bump tested or calibrated,
and if the option to bypass bump testing or calibration is not available, you should shut off the
instrument and follow the procedure outlined here if you want to change the Policy
Enforcement settings:
1. Use a USB cable to connect the AreaRAE to a computer running Safety Suite Device
Configurator (SSDC).
2. Enter Diagnostic Mode on the AreaRAE Plus/Pro (with the instrument turned off, press
and hold [Y/+] and [MODE] until it starts up.
3. After startup, enter the password when prompted (default is 0000”) and press [MODE].
4. Press [N/-] repeatedly until you see the Enter Communications Mode? screen.
5. Press [Y/+] to enter Communications Mode.
6. Safety Suite Device Configurator (SSDC).
7. Input the username and password (the default username is administrator and the
default password is "Default123").
8. Click LOGIN
9. Click on the instrument’s serial number.
10. Click on the Settings” tab.
11. Scroll down to Policy Settings.
12. Deselect Policy Enforcement features you do not wish to use.
13. Click Save”
14. When the upload is done, exit Safety Suite Device Configurator (SSDC).
15. Press [Y/+] on the AreaRAE Plus/Pro to exit Communication Mode.
AreaRAE 87 User Manual

Table of Contents

Other manuals for Honeywell AreaRAE Plus

Related product manuals