External Battery Charger
You can charge a spare Li-Ion battery with the optional External Li-Ion Battery Charger For
AreaRAE (P/N: W01-3005-000).
Once the Charger is attached to the battery, plug in the AC adapter’s barrel and the plug the
power cord into an AC source.
WARNING
To reduce the risk of ignition of hazardous atmospheres, recharge, remove or replace the
battery only in an area known to be non-hazardous! Do not mix old and new batteries or
batteries from different manufacturers.
External Battery Charger LED Indicators
LEDs on the External Battery Charger indicate the following:
LED Indication
Solid red NTC fault, bad battery, battery standby
Blinking
red
Charging, but the capacity is too low to power the instrument (less than 10%)
Blinking
green
Charging. The capacity is sufficient to power the instrument (more than 10%, but less
than or equal to 99%)
Solid green The LCD shows that the battery is fully charged when any button is pressed.