Note: When a pump alarm occurs and the AreaRAE Plus/Pro is monitored by Safety Suite Device
Configurator (SSDC), the PID’s reading in SSDC shows - - -”.
Calibration Status
The instrument displays this icon next to the sensor that requires calibration:
Calibration is required (and indicated by this icon) if:
l The lamp type has been changed.
l The sensor module has been replaced with one whose calibration is overdue.
l The defined period of time between calibrations has been exceeded.
l If you have changed the calibration gas type without recalibrating the instrument.
l The sensor has failed a previous calibration.
Bump Status
The instrument displays this icon next to the sensor that requires bump test:
A bump test is required (and indicated by this icon) if:
l The defined period of time between bump tests has been exceeded (bump test overdue).
l The sensor has failed a previous bump test.
l The sensor(s) should be challenged on a periodic basis.