If you make a mistake, you can cycle through the digits by pressing [N/-] and then using [Y/+] to
change the number in each position.
Note: The default password is 0000.
Note: The password screen appears when you enter the Programming Mode the first time after
turning the instrument on in Basic Mode, and each time after that.
Once you enter Programming Mode, the Calibration menu is highlighted. You cannot access
other menu items, and you can only perform the following types of calibration:
l Fresh Air
l Multi Sensor Span
l Single Sensor Zero
l Single Sensor Span
The following require password access in Basic Mode or can be directly accessed in Advanced
Mode.
l Multi Sensor Bump
l Single Sensor Bump
l Cal. Reference
l Change Cal. Gas
l Multi Cal Select
l Change Span Value
To enter a menu and view or edit parameters in its submenus, press [Y/+].