To perform a bump test on an individual sensor, follow these steps:
1. At the Calibration Menu, select Single Sensor Bump.”
2. Scroll down the list using [N/-], and then press [Y/+] to select a sensor to bump test.
3. Install the T” tube and connect it to a source of calibration gas.
4. Verify that the displayed calibration value meets the concentration specified on the gas
cylinder.
5. Start the flow of calibration gas.
Note: Dotted line indicates automatic progression.
6. Press [Y/+] to start calibrating or wait for the bump test to start automatically.
7. A countdown screen appears. You can abort the bump test at any time during the
countdown by pressing [N/-].
8. If the bump test is not aborted, the display shows the sensor names and tells you whether
the bump test passed or failed, followed by the sensor readings.
AreaRAE 59 User Manual