EasyManua.ls Logo

Honeywell BW MaxXT II - Cal IR Lock Configuration; Cal IR Lock Enabled Status and Implications

Honeywell BW MaxXT II
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
The sensor(s) must be calibrated to continue and enter normal operation. For complete
instructions, refer to See "Calibration" on page70 for more information..
Note: If the calibration is unsuccessful for any of the sensors, the detector deactivates.
Force Calibration Disabled: If disabled, the CAL DUE NOW screen displays. Press to
acknowledge the warning and enter normal operation.
The detector is shipped with the Force Calibration option disabled.
Note: If overdue for calibration, BW recommends the sensor(s) be calibrated immediately.
Cal IR Lock
Note: The auto-zero function is not affected if the CAL IR Lock option is enabled. The detector
still performs the auto-zero function (if enabled).
BW MaxXT II 31 User Manual

Other manuals for Honeywell BW MaxXT II

Related product manuals