Calibration
Part No.: 44200004 - Revision 2 SVP Controller Operation Manual 4-17
After selecting Enter Average Volume following screen appears
Enter Repeatability
0.00200 %
7 8 9 Backspace
4 5 6 Confirm
1 2 3
0 - .
Enter value Stop
Selecting “Abort Calibration” will stop the calibration process. This action
will stop the motor if in running state, bypass value (SV2) will be turned ON
and draw valve (SV1) will be turned OFF. If the Prover is in sweep mode
with the piston moving downstream no action will be taken for upstream
and downstream sensor inputs. Calibration log entry will not be made.
In calibration mode alarms will be generated as per the configuration, and
the run permissive signal from the flow computer or the run prover test
signal will be ignored.
The operating mode of the prover will be saved in NVM so that the mode
will be restored if power is turned off while in Prover Calibration mode.
To exit calibration mode navigate to the Exit Calibration Mode menu and
press OK.
A copy of Table 2 can be used to enter water draw data.