Procedural Considerations
Page 52 34-ST-25-20 MC Toolkit User Manual Release 3
9/06
Honeywell HART Calibration Zero Trim
Menu
Honeywell ST 3000
(Note - Zero Trim is
available.)
Honeywell STT 3000
(Note - no Zero Trim.)
Zero Trim
NOTE:
This procedure will
change LRV and URV
settings.
Select Zero Trim. The first
of a series of Pop-Up
messages appears.
To acknowledge the
message, select the Yes
button; another message
appears.
Apply the zero-reference
input source to the sensor.
Select the
OK button, and
wait for this message:
Select the
OK button; the
following message should
appear.