MC Toolkit Software with Honeywell HART Transmitters  
Release 3   34-ST-25-20   MC Toolkit User Manual  59 
9/06   
Honeywell HART Calibration - Apply Values 
Apply 
Values: 
LRV 
 
Select the OK button.  The display at right 
appears.   
The Current Applied Process Value field 
shows the value of the Process Value.  A 
new sample of the input level is displayed 
each time the user selects the Read New 
Value button. 
Adjust the process variable to the desired 
value while repeatedly selecting the Read 
New Value button to monitor and verify 
the input value. 
 
 
(Write 
input value 
as LRV.)
 
When the Process Variable 
is stabilized at the desired 
input level, select the Set 
New Value button.   
The popup at right appears.  
Select the OK button to 
write the input value as the 
LRV calibration value; and 
the popup message box will 
disappear. 
 
 
 
 
(Adjust 
and View 
Process 
Variable 
input value 
at URV.)
 
Select the 20 mA button.  
The popup at right appears. 
 
This display at right is the 
same one used for applying 
PV input as LRV value. 
Adjust the PV input to the 
desired URV level while 
using the Read New 
Value button for 
monitoring.