Mini-AT User Guide
Item
No. Item Name Description
————————————————————————————————————————
106 TEMP OUT RANGE If item 106 is programmed into item 072, 073 or 074, this item
(106) indicates if a high or low temperature alarm was
generated. During a wake-cycle,
if the measurement for item
026 (GAS TEMPERATURE) is a value less than the value at item
027 (TEMP LO LIM) or a value greater than the value at item 028
(TEMP HI LIM)
, an alarm is initiated and is indicated by
placing dots on the LCD and “11111111” at item 106.
“00000000” at item 106 indicates there is no Temperature
Alarm. The alarm indicators will remain active until manually
cleared (by software) or automatically when RBX (item 165) is
enabled. Also see items 144 and 146.
Default = 00000000
108 ALARMS OUTPUT This item
(108) displays “11111111” to indicate that a standard
alarm (069-071, 099-104, 143-146, 163 or 222) or a selected
optional alarm (105 or 106) has become active, and that an alarm
pulse was transmitted out the Alarm Channel.
“00000000” at
item 108 indicates no alarms became active since the last
‘Clear Alarms’.
Default = 00000000
109 FIXED PRESS FACT
Code (0 or 1) that selects Live or Fixed Factor Pressure. When
enabled for Fixed Factor, the Pressure Factor entered at item 044
will be used for pressure correction, instead of a Live pressure
factor calculation based on the input from the pressure transducer
(item 008). NOTE: if item 109 is “Fixed”, a value must be entered
at item 044. It may also be necessary to enter a value at item 008
(Gas Pressure) if Supercompressibility (item 110) is “Live”.
Select: 0 - No Fixed Factor (Live) Default
1 - Fixed Factor
110 FIXED SUPER FACT
Code (0 or 1) that selects Live or Fixed Supercompressibility.
When enabled for Fixed Super, the Unsquared Super Factor
entered at Item 047 will be used for Supercompressibility
Correction, instead of a Live Super calculation. NOTE: if item 110
is “Fixed”, a value must be entered at item 047. Entering a fixed
Super value of “1.0000” will essentially
disable the
Supercompressibility function.
Select: 0 - No Fixed Factor (Live) Default
1 - Fixed Factor