RAEGuard 3 User Manual
34
4. Select the “√switch to enter the “Testscreen as shown below.
5. Select the “Bump Testand use the “√switch, at the same time the detector
enters “Inhibitmode. If it is an oxygen sensor, the prompts are as follows:
6. If it's not an oxygen sensor, the menu goes to "BumValue Set". Normally, the
default bump gas concentration is 50% of the full scale, and the range can
be set from 30% to 70% F.S.
7. Use the switch to edit the bump test gas concentration you need within
the settable range according to the actual conditions on site. Use the switch
“▲” to increment or decrement; Use the switch to move to the next
character; Select the “√switch to complete the setting.
8. Use the “√” switch to save the settings and display as follows