12 Menu’s and Advanced Configuration
RAEGuard 3 offers users a variety of operating modes, providing different
management permissions for operators in different functions.
After entering the menu below, the 4-20 mA current loop is locked in the
"Inhibit" current, and the default "Inhibit" current is 2 mA.
Menu Item
4-20 mA current output
1
Warm/Start up
Inhibit current output
2
Sensor Replacement
Inhibit current output
3
Alarm & fault simulation
Inhibit current output
4
4 mA output calibration
4 mA
5
20 mA output calibration
20 mA
6
Forced relay
Inhibit current output
7
Forced current output
Output corresponding set value
8
Bump test
Inhibit current output
9
Zero & Span calibration
Inhibit current output
10
Others
Real-time detection current
Put the magnet
over the switch
“√” for more than
3s to enter
Configuration
Mode.
Select the
switch “√” to
enter View
Mode