RAEGuard 3 User Manual
43
Check the main board, or replace it
Check the sensor module, or replace it, or
contact the factory
Communication
failure between
sensor and
transmitter
Check the cable and safety barrier, or
replace them, or contact the factory.
Perform "Date Setting" and "Time Setting" in
the menu at the same time.
Calibration expiration
reminder
Please recalibrate the detector
Perform zero calibration on the detector or
replace the sensor.
Sensor life expiration
warning
Please replace the sensor and calibrate it.
After confirming that there is no leakage on
site or the alarm is released, please click the
“√” button to “reset” the latch.
After confirming that there is no leakage on
site or the alarm is released, please click the
“√” button to “reset” the latch.
Indicating that the gas cumulative
concentration in 8 hours exceeds the
standard
Indicating that the gas cumulative
concentration in 15 minutes exceeds the
standard
Indicating that the gas concentration
exceeds the range, and the detector returns
to the normal detection mode after the site is
safe.
Indicating that the gas concentration on site
exceeds the underreportting set value.
Indicating that the on-site gas concentration
exceeds the high alarm setting value.
Note: When the detector exceeds the Full-Scale Range, the normal detection
mode is displayed after the on-site gas concentration returns to the normal
range. But for Catalytic Bead sensor, its high working current, long time working
under over-range conditions will lead to sensor damage. The detector will
detect the gas concentration in the environment every 4 minutes until the
ambient gas returns to the normal range, and the detector returns to the normal
detection mode.