Password settings
o Low-level password setting111
o Advanced password settings: 888
Communication settings
o Address: 001
o Baud rate:
4800;9600;19200;38400;57600
o HART address:00
Output crrent lvel
o Fault mode
1.0 mA;2.0 mA;3.0 mA
o Inhibit mode
1.0 mA;2.0 mA;3.0 mA
o Warning mode
1.0 mA;2.0 mA;3.0 mA
o Out of range
21.0 mA;22.0 mA;23.0 mA
4-20 mA Output calibration
o 4 mA output calibration
o 20 mA output calibration
Restore factory settings
Note: The blackened font in the table above is the factory default value.
Zero & Standard Gas Calibration: See section 8 for details.
Calibration time interval: This menu is to remind users to calibrate regularly.
The default value is 90 days. According to the actual situation, users can set
it in 0-360. If 0 is set, the reminder function will be turned off; if other values
are set, the detector will display and the LCD backlight will be always on
when the calibration interval time is reached.
Bump test: See section 8 for details.
Forced current output: 4 mA by default. It can force 4-20 mA current loop to
output the set current value between 0-25, and the increment and decrease
step is 0.5 mA. At the same time, the detector FAULT LED is always on,
indicating the “Inhibitmode.