Page 124 SMV800 Modbus User's Guide Revision 1.0
ts_sensor_over_
temperature
The Process Temperature
input exceeds
the Temperature Upper
Range Limit
(URL) as determined by
the configured
Sensor Type.
Check the process
temperature. If the
process temperature
exceeds the range of
the current sensor type,
either correct the
process to an in-range
temperature or switch to a
different sensor type which
is ranged for the expected
process temperature
range.
The temperature
calibration correcs on LRV,
URV or both is in excess.
Perform Reset Corrects to
reset the User calibration
to factory default. If
required, fllow the
calibration procedure to
repeat the temperature
calibration.
This is acknowledgement
status for reset corrects
performed by user
Temperature module input
bad. Input out of range or
may be open
Use matching sensor
suitable for the process
range. Reset the device,
replace the Temp Module if
issue persists
Temperature Module CJ
measurement is bad
Verify that the CJ sensor
within the Temp sensor
module is not outside of
the operating temp limits (-
40 to 85 degC). Reset the
device, replace the Temp
Module if issue persists
Fault is set if Temperature
module
• input failue faulty sensor,
• out of range
• input open
Replace the faulty sensor if
that is the reason. Select
suitable sensor type for the
process. Reset the device,
replace the Temperature
Module if issue persists
Temperature module
Supply is Low
Check supply voltage is
within specification. If all
are within specification,
replace Temperature
module