Revision 1.0 SMV800 Modbus User's Guide Page 23
2.3.5. Flow Configuration
This tab provides the detail information about the Flow parameters.
Refer the below table for Flow Configuration:
Flow Configuration parameters
Key: Plain = Read only Bold = Configurable Bold underline = Method Bold italic = Table or graph
Flow Config
Flow
The current value of the calculated Flow
Flow LRV
The Lower Range Value for the Flow input
(which represents 0% output) in user selected
engineering units. This value may be
configured to any value within the range Flow
LTL to Flow UTL.
Flow URV
The Upper Range Value for the Flow input
(which represents 100% output) in user
selected engineering units. This value may be
configured to any value within the range Flow
LTL to Flow UTL.
Flow Damping
Damping value for the Flow output. Entries
may be any value from 0.00 to 32.00 seconds.
The upper limit for Flow damping is 100.
Flow URL
The Upper Range Limit for the Flow input
(editable)
Flow LRL
The Lower Range Limit for the Flow input